Reason stated in the source
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Source facts
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2019-03-20
- Root cause
- Device Design
- Terminated
- 2020-10-05
- Distribution
- US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
- Recall number
- Z-1258-2019
- Recall status
- Terminated
- CDRH record ID
- 171912
- FDA record created
- 2019-05-03
- Recall event number
- 82582