Z-1258-2023: Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measureme

Recall record Z-1258-2023 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1258-2023
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Mar 15, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L) (103% positive bias).

Source facts

Action
Siemens issued to US Customers Urgent Medical Device Correction (DC-23-03.A.US) via FedEx beginning on 01/30/2023. OUS Customers sent Urgent Field Safety Notice (DC-23-03.A.OUS). Letter states reason for recall, health risk and action to take: Discontinue use and discard the lots listed in Table 1 and complete the attached Product Replacement form for no-charge product replacement. Please note, lot GA2286 is expired. Review your inventory to determine possible replacement needs and provide information to Siemens for reporting to the Authorities. Please review this letter with your Medical Director and the decision on a review of previously generated results. If you receive any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Remote Services Center or your local Siemens Technical Support Representative. Complete and return the Field Correction Effectiveness Check/Product Replacement Form attached to this letter within 30 days to Siemens Healthineers for reporting to the Authorities. Questions, contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers Technical Support Representative.
Initiated
2023-01-30
Root cause
Under Investigation by firm
Distribution
Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile Colombia Country India Japan New Zealand Nicaragua Saudi¿Arabia¿¿¿ South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Taiwan
Recall number
Z-1258-2023
Recall status
Open, Classified
CDRH record ID
198493
FDA classified
2023-03-15
Recall event number
91677

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗