Z-1259-06: Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500

Recall record Z-1259-06 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1259-06
Company in FDA record
Beckman Coulter Inc
FDA classification date
Jul 26, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Beckman Coulter has determined that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results.

Source facts

Action
A Product Corrective Action letter was sent via US mail the week of May 22, 2006 to all FC500 customers. They were informed that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results. The letter includes the sequence in which the issue occurrs: 1) During the acquistion of a tube, the Pause and Rotate Button is selected. 2) A system error is generated during the Pause and Rotate routine (The error message table may be viewed by going to the CXP Help Menu, select the CXP Help Menu Item, then to Section 13.5). 3) An error message appears in the Cytometer Status, an audible alarm sounds, the run aborts, the carousel ejects, and the system automatically transitions from Verification to Run Initialization and then to Awaiting Sample 4) The Worklist is continued or a new Worklist is loaded and started. 5) After the Worklist is continued, or loaded and started, during the acquistion of a tube, the Pause and Rotate Button is selected. 6) The carousel Pauses but does not Rotate. 7) When the Start/Continue or Restart button is selected, the carousel may move to the wrong tube and aspirate sample resulting in a sample misidentification. If no system errors are generated during a Pause and Rotate routine, the system will function correctly. The letter also provides steps for immediate user preventive action if a system error is generated during the Pause and Rotate Routine to remedy the issue: 1) Exit CXP Software. 2) Power off the Cytometer using the FC OFF utility 3) Turn the Cytometer back on. 4) Recovery should be complete and the laboratory may resume normal usage of the system. Customers are asked to share the information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation and to complete and return the enclosed Fax Response Form for the firm''s records.
Initiated
2006-05-22
Root cause
Other
Terminated
2012-04-26
Distribution
Nationwide and Canada
Recall number
Z-1259-06
Recall status
Terminated
CDRH record ID
46535
FDA classified
2006-07-26
Recall event number
35643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗