Z-1259-2022: illumina REF DX-410-1001 Model: MiSeq Dx illumina REF 15036706 Model: MiSeq Dx illumina REF 20014053 Mo

Recall record Z-1259-2022 for Illumina, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1259-2022
Company in FDA record
Illumina, Inc.
FDA record creation date
Jun 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

cybersecurity vulnerability

Source facts

Action
On 05/03/2022, Illumina sent via email an "Urgent Medical Device Recall" Letter to customer informing them the cybersecurity vulnerability impacting the NextSeq 550 Dx and MiSeq Dx instruments with software versions 1.3 to 3.1 of the Local Run Manager (LRM) software installed. Customers are also being informed that Illumina has developed a software patch to protect against the exploitation of the vulnerability. Installation of the software patch will block remote access to the LRM web User Interface (web UI). For NextSeq 550Dx and MiSeq Dx instruments connected to the internet, the software patch is available for immediate download through a secure portal. Illumina is currently providing the website address to impacted customers and regulatory authorities only. Illumina recommends that customers download and install the software patch on their instrument(s) promptly. For affected NextSeq 550Dx and MiSeq Dx instruments not connected to the internet, Illumina has developed other options for the installation of the software patch. Impacted customers should contact Tech Support promptly to obtain information about these options. NextSeq 550Dx and MiSeq Dx instruments that have dual boot mode, the customers must install the software patch separately in each mode (Dx mode and RUO mode) of each instrument. It is also important for the customer to install the software patch on any stand-alone instances of the off-instrument LRM for RUO mode on the Dx instruments. Illumina will ask the customer to complete the Urgent Medical Device Recall verification form to confirm receipt of this notification and verify the patch was installed. We will ask the customer to return the form to Technical Support. For questions, email - techsupport@illumina.com
Initiated
2022-05-03
Root cause
Software in the Use Environment
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, and WV. The countries of Algeria, Australia, Austria, Belgium, Bulgaria, Bosnia/Herzegovina, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Guatemala, Hungary, Ireland, Italy, Japan, Latvia, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, New Zealand, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1259-2022
Recall status
Completed
CDRH record ID
193602
FDA record created
2022-06-13
Recall event number
90160

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗