Open original FDA record ↗Recall record
- FDA identifier
- Z-1260-2009
- Company in FDA record
- Respironics
- FDA classification date
- Jun 17, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Malfunction of the J3 connection between the Heater Plate Main printed circuit assembly (PCA) and the heater plate may have an intermittent connection which may lead to excessive heat build-up on the humidifier PCA.
Source facts
- Action
- A Service Communication dated February 5, 2009 was provided to all Respironics customers that are approved to service Respironics devices. The communication described the affected device and actions required. Further questions may be addressed by phone at 1-800-345-6443, and by email at service@respironics.com or clinical@respironics.com.
- Initiated
- 2009-02-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-07-05
- Distribution
- Worldwide Distribution -- US including states of OR, OH, KY, TX, MO, AZ, IN and NV and countries of United Arab Emirates, Bosnia, Belgium, Bulgaria, Bahrain, Canada, Czech Republic, Denmark, Egypt, Spain, Finland, Greece, Croatia, Hungary, Ireland, Israel, Jordan, Kuwait, Lebanon, Latvia, Macedonia, Netherlands, Norway, Oman, Poland, Portugal, Romania, Qatar, Russia, Slovenia, Turkey, Serbia and South Africa.
- Recall number
- Z-1260-2009
- Recall status
- Terminated
- CDRH record ID
- 81028
- FDA classified
- 2009-06-17
- Recall event number
- 51699