Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
Source facts
Action
The recalling firm initiated this recall with a letter to send to their customers, via Registered Mail, on 5/1/2006. The letter requests that the customers identify and discard the potentially defective cartridges from Lot #R0299. The letter is accompanied by a recall response page to be completed by the customer and faxed or mailed back to the Radiometer so that the customer can receive an equal number of replacement cartridges free-of-charge.
Initiated
2006-05-01
Root cause
Other
Terminated
2012-02-24
Distribution
Nationwide.
Recall number
Z-1261-06
Recall status
Terminated
CDRH record ID
46638
FDA classified
2006-07-27
Recall event number
35698