Open original FDA record ↗Recall record
- FDA identifier
- Z-1261-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Mar 28, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function, disc capacity for free space decreases faster than usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged "Not to be archived" must be manually prevented from being deleted by either clinical administrator or service engineer.
Source facts
- Action
- Customers received letters with information related to the issue and regarding the upgrade to the software version VB20A_HF04. The software was released December 22, 2017 and is available for install.
- Initiated
- 2018-01-10
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2018-10-02
- Distribution
- Medical device software which needs to be installed.
- Recall number
- Z-1261-2018
- Recall status
- Terminated
- CDRH record ID
- 162538
- FDA record created
- 2018-03-28
- Recall event number
- 79513