Skip to content

Public FDA record

Z-1261-2023: TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/

Recall record Z-1261-2023 for Limacorporate S.p.A. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1261-2023
Company in FDA record
Limacorporate S.p.A
FDA classification date
Mar 16, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.

Source facts

Action
On or about 01/25/2023, the firm sent a "VOLUNTARY CORRECTION" Letter via email to customers informing them that the firm has become aware that it incorrectly included low constraint Tema axle small components (not cleared for marketing in U.S.) with its high constraint humeral bodies system. According to the firm's records, 3 low constraint components were implanted between 12/10/2021 to 07/05/2022. Customers are being instructed to: LimaCorporate suggests to Surgeons and Hospitals where an affected device listed in Table 1 was implanted in combination with high constraint humeral bodies components belonging to the product family 1550.15.0XX to yearly follow-up patients with radiographs and varus/valgus range-of-motion evaluation to assess for poly bushing wear. If the follow-up monitoring after 24 months from the implant date shows uneventful and no abnormality, surgeons are invited to do patient follow-up according to the standard protocols, except in case of pain subsequently experienced by the patient. In addition, customers will be instructed by the firm's U.S Regulatory Affairs Associate to post the letter in or near implanted patients' records to continuously remind surgeons and hospitals to monitor their implant patients regarding this Voluntary Correction. For questions or further assistance, contact the firm via email at medicalcomplaints@limacorporate.com.
Initiated
2023-01-25
Root cause
Employee error
Distribution
U.S. Distribution to: TX
Recall number
Z-1261-2023
Recall status
Open, Classified
CDRH record ID
198354
FDA classified
2023-03-16
Recall event number
91613

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗