- Recall classification
- Class II
Reason stated in the source
The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient.
Source facts
- Action
- On February 2, 2018, the firm issued an Urgent Medical Device Correction notification to their distributors to make them aware of the device correction. The firm requested that distributors further disseminate an attached recall letter to consignees. The communication addressed to consignees advised of the issue affecting the device. Consignees were asked to inform users that the device is for armboard use only, and not to be used with the product system's leg support. The notice informed consignees that the firm would provide new units upon completion of development along with revised Instructions for Use. Customers with questions concerning this Urgent Medical Device Correction were referred to Allen Medical Customer Service: (800) 433-5774, 8:30AM-5PM (EST)
- Initiated
- 2018-02-21
- Root cause
- Error in labeling
- Terminated
- 2020-04-06
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia, France, Saudi Arabia, Germany, and France.
- Recall number
- Z-1262-2018
- Recall status
- Terminated
- CDRH record ID
- 162773
- FDA record created
- 2018-03-28
- Recall event number
- 79613