Open original FDA record ↗Recall record
- FDA identifier
- Z-1263-2009
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- May 14, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery.
Source facts
- Action
- An "Urgent - Medical Device Recall" Letter dated January 2009 was issued to consignees via certified mail. The letter described the issue, specific conditions, background information, and precautions for customers. An additional letter dated January 22, 2009 was included with the "Urgent - Medical Device Recall" letter which instructed consignees to inform an enduser of the contents of the notification letter because some of the consignees are home care device/service providers. When the new hardware/software solution is available, affected devices will be updated free of charge. Direct questions about the notification letter to Drager Medical, Inc. by calling 1-800-543-5047 (press "1" at the prompt and then press "2349#"). Direct questions regarding updating your device when the new software is available, contact Drager Service Technical Support at 1-800-543-5047 (press "4" at the prompt).
- Initiated
- 2009-01-21
- Root cause
- Software design
- Terminated
- 2010-01-05
- Distribution
- Nationwide Distribution -- states of CA, FL, GA, HI, IL, LA, MA, MN, OH, and TX.
- Recall number
- Z-1263-2009
- Recall status
- Terminated
- CDRH record ID
- 80299
- FDA classified
- 2009-05-14
- Recall event number
- 51447