Skip to content

Public FDA record

Z-1263-2024: NO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM

Recall record Z-1263-2024 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1263-2024
Company in FDA record
Howmedica Osteonics Corp.
FDA classification date
Mar 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential packaging breaches of inner blister and outer sterile blister.

Source facts

Action
URGENT MEDICAL DEVICE RECALL notifications dated 02/06/2024 were distributed to consignees by mail. Consignees were asked to forward the provided recall notice to all individuals within the organization who need to be made aware and organizations who have consigned product, immediately check all stock and operating room storage for affected devices, quarantine and discontinue use of recalled devices, complete and return the provided Business Reply Form by email to strykerortho4004@sedgwick.com or by fax to 855-800-1093. Consignees are to contact their local sales office or Stryker Sales Representative for product replacement and inventory questions. Recalled devices are to be returned to Stryker Orthopaedics/PFA Product Returns, Attn: Distribution Inventory Team, 325 Corporate Drive, Dock M-East, Mahwah, NJ 07430; include "Ref PFA 3524019" on shipment. Consignees with any questions are to call the firm at 201-831-5000 or by email at SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com. An URGENT MEDICAL DEVICE RECALL UPDATE, dated 3/18/24, was sent to consignees.
Initiated
2024-02-06
Root cause
Packaging
Distribution
Domestic: Nationwide Distribution; Foreign: Australia, Canada, Chile, Colombia, Korea, & Japan.
Recall number
Z-1263-2024
Recall status
Open, Classified
CDRH record ID
206157
FDA classified
2024-03-06
Recall event number
94026

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗