Z-1264-2021: Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100

Recall record Z-1264-2021 for Imactis. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1264-2021
Company in FDA record
Imactis
FDA record creation date
Mar 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory

Source facts

Action
Imactis notified users by visit on 2/11/221 and issued Urgent Medical Device Recall on 2/25/21 via email to consignees. Letter states reason for recall, health risk and action to take: Customers are requested to quarantine any product from the affected NaviKit" lot numbers and return them to Imactis Affected NaviKit lot numbers will be quarantined and removed from the market. New compliant product will be sent to customers to replace the removed lot numbers. As the defect is visually detectable, product will be inspected internally according to an approved procedure and any defective needle holders will be sorted and replaced with compliant ones. Refurbished NaviKit" product will not be sold to customers. Your IMACTIS representative for questions: Robert Hornak (VP Sales, North America) - 205-542-1486 (Cell) - 205-542-1486 (cell) Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time
Initiated
2021-02-11
Root cause
Process control
Terminated
2023-10-05
Distribution
NJ, NY, WI
Recall number
Z-1264-2021
Recall status
Terminated
CDRH record ID
185949
FDA record created
2021-03-16
Recall event number
87391

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗