Z-1265-2021: Incisive CT Computed Tomography X-ray System Model: 728143

Recall record Z-1265-2021 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1265-2021
Company in FDA record
Philips North America Llc
FDA record creation date
Mar 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees

Source facts

Action
Philips notified US consignee on 2/22/2021via email including the Field Safety Notice informing of the reason for recall, health risk and action to take: stop using the system until a Philips Field Service engineer arrives on site to implement the correction to lock the gantry tilt function (FCO72800745). After the FCO72800745 implementation, the device can continue to operate without tilt function until the final correction (FCO72800742) is implemented to replace the nonconforming actuators which will remedy the issue and return device to normal use. A Philips Service Engineer will contact you to schedule the Field Actions. Both of the Field Actions FCO72800745 & FCO72800742 will be free of charge.
Initiated
2021-02-22
Root cause
Process control
Terminated
2023-02-09
Distribution
US Virgin Island Foreign: Australia Bolivia China Germany Hungary Indonesia Italy Japan Korea Korea, Republic of Latvia Palestine Philippines Romania Russian Federation Senegal Thailand Uzbekistan
Recall number
Z-1265-2021
Recall status
Terminated
CDRH record ID
186005
FDA record created
2021-03-16
Recall event number
87427

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗