Skip to content

Public FDA record

Z-1266-06: Terumo Advanced Perfusion System 4 inch diameter Roller Pump; Model 801040.

Recall record Z-1266-06 for Terumo Cardiovascular Systems Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1266-06
Company in FDA record
Terumo Cardiovascular Systems Corp
FDA classification date
Jul 27, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The pump may stop during the autodose delivery without completion of the dose delivery to the patient.

Source facts

Action
Service technicians were notified on 9/30/05 to upgrade the software on these units by March 31, 2006 as follows: pump software to version 1.20, power manager software to version 1.20 and monitor software to version 2.2.1. U.S. software has all been upgraded and international upgrades are ongoing as of 7/7/06.
Initiated
2005-09-01
Root cause
Other
Terminated
2008-06-11
Distribution
Nationwide, and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand, Turkey and United Arab Emirates.
Recall number
Z-1266-06
Recall status
Terminated
CDRH record ID
43983
FDA classified
2006-07-27
Recall event number
34488

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗