Open original FDA record ↗Recall record
- FDA identifier
- Z-1266-2009
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 28, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configured to archive image data on a network file system (NFS) connected via a connected archive (Mount Point Solution).
Source facts
- Action
- A Customer Safety Advisory Notice, dated 02/09/2009 was sent to affected customers via Update Instruction IM013/09S. The letter informs customers of when the malfunction occurs and the potential issue, provides instructions to avoid its occurrence, and how the issue will be resolved. Customers are to ensure that the Safety Advisory is placed in the system's instructions for use and contact their SIEMENS Uptime Service Senter if they have any questions.
- Initiated
- 2009-02-09
- Root cause
- Software design
- Terminated
- 2009-09-23
- Distribution
- Nationwide Distribution -- including states of AL, CA, CT, FL, LA, MI, MO, NC, NJ, OH, PA, TN, and WI.
- Recall number
- Z-1266-2009
- Recall status
- Terminated
- CDRH record ID
- 80715
- FDA classified
- 2009-04-28
- Recall event number
- 51563