Z-1266-2016: The S.C.O.R.E.S. Unit (self contained operating room equipment sterilization unit)

Recall record Z-1266-2016 for PMBS, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1266-2016
Company in FDA record
PMBS, LLC
FDA record creation date
Mar 25, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Water retention issues. Units with filters that remain wet after the sterilization may compromised sterility.

Source facts

Action
On January 13, 2016 PMBS, LLC sent letters to their customers informing them of the correction. Per product's instructions for use, the SCORES cabinet is to be run for a minimun of a 4 minute sterilization cycle, followed by a 30 munites dry time followed by a 30 minutes door crack time. not all facilities generate completely dry SCORES filters following these instructions. It is imperative that the filters be checked in the Operation Room prior to useing the surgical instruments sterilised within the cabinet. If the filters are wet, do not use the contents. In the letter is a IFU addendum including instructions for running the equiptment with extended cycles. For more information please contact Ms. Michele Mauzerall, CEO at 908-300-7093.
Initiated
2016-01-13
Root cause
Device Design
Terminated
2017-09-27
Distribution
FL, GA, MN, NY, OH, OR SC, and TN
Recall number
Z-1266-2016
Recall status
Terminated
CDRH record ID
143282
FDA record created
2016-03-25
Recall event number
73157

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗