Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard
Source facts
Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
Healthcare practitioners are instructed to courier their TLC-2000 to Theralase. The TLC-2000 will be updated to reduce EMI to meet IEC 60601-1-2 standard (4th edition).
Initiated
2019-09-26
Root cause
Process control
Terminated
2020-12-03
Distribution
Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Recall number
Z-1266-2020
Recall status
Terminated
CDRH record ID
179532
FDA classified
2020-02-13
Recall event number
84846