Possibility of image artifacts during data acquisition when using Adaptive Cardio Sequence, Turbo Flash, and Head modes. The update improves system start-up behavior and resuming, ECG handling with visual feedback of correct placement of ECG leads and contact quality, and other safety related issues.
Source facts
Action
Siemens sent a Customer Advisory Notice dated February 6, 2015, to all customers via USPS Certified Mail. The notice identified the affected product, problem and actions to be taken.
Initiated
2015-02-06
Root cause
Software design
Terminated
2015-08-31
Distribution
US Nationwide Distribution in the states of PA, MA, MN, NY, SC, MD, NY, IL, OH, TX and KY.
Recall number
Z-1267-2015
Recall status
Terminated
CDRH record ID
134030
FDA classified
2015-03-12
Recall event number
70556