Skip to content

Public FDA record

Z-1267-2024: ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b

Recall record Z-1267-2024 for ARROW INTERNATIONAL Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1267-2024
Company in FDA record
ARROW INTERNATIONAL Inc.
FDA classification date
Mar 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.

Source facts

Action
Teleflex/ARROW INTERNATIONAL issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 02/12/2024 via letter delivered via FedEx. The notice explained the issue, potential risk, and requested the following actions: Identify affected product in inventory, cease use, and quarantine pending return. If affected product is found, mark the applicable checkbox on the Acknowledgement Form and fax it to 1-855-419-8507, including Attn: Customer Service or using the e-mail address recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. If no product is inventory or on hand, mark the applicable checkbox on the Acknowledgement Form and return the form to Teleflex at the contact details provided. Distributors: Provide this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. Immediately check your inventory for impacted product. Cease use and distribution of impacted product and immediately quarantine the affected product. You may then return all product in scope. Should you require further information or support concerning this issue, please contact Customer Service via email: recalls@teleflex.com, phone: 1-866-396-2111, or FAX: 1-855-419-8507 using the details provided below. Customer Service hours of operation are 8am 7pm EST.
Initiated
2024-02-12
Root cause
Process control
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU, SA, SE, SG, SI, TH, TW, and UK.
Recall number
Z-1267-2024
Recall status
Open, Classified
CDRH record ID
206174
FDA classified
2024-03-21
Recall event number
94033

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗