Open original FDA record ↗Recall record
- FDA identifier
- Z-1268-06
- Company in FDA record
- Varian Medical Systems
- FDA classification date
- Jul 27, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment data is entered manually.
Source facts
- Action
- The recalling firm notified end users by letter on 05/01/06 and advised users of possible communication error between afterloader devices and software. The notification also advised that the control program for the devices has the potential to lead to a misadministration of treatment in cases where data is manually entered by the user. The firm additionally advised that user may continue to use the afterloader devices if following the listed cautionary statements. A software upgrade is planned to be implemented at the user site''s next regularly scheduled radiation source exchange. A response form is enclosed with the notification and users are requested to return the customer receipt verification via fax.
- Initiated
- 2006-05-01
- Root cause
- Other
- Terminated
- 2007-05-18
- Distribution
- Nationwide and worldwide.
- Recall number
- Z-1268-06
- Recall status
- Terminated
- CDRH record ID
- 46634
- FDA classified
- 2006-07-27
- Recall event number
- 35694