Reason stated in the source
Automatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes diluted patient samples to under recover by approximately 50%.
Source facts
- Action
- The firm, Siemens Healthcare Diagnostics, sent an "Urgent Device Recall Notice" on February 16, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to discontinue using the automatic and system-performed dilution features for Vitamin D; to discontinue the use of ADVIA Centaur VitD Diluent, 2-pack (REF 10494100); to manually dilute samples with values greater than the assay range of 150 ng/mL (375 nmoI/L), using the ADVIA Centaur VitD Diluent, 1 bottle (REF 10632114) as stated in the Vitamin D assay Instructions For Use (IFU); to review the contents of the recall notice with their Laboratory Director and retest the previous sample results generated using system- performed dilutions, and complete and return the confirmation fax-back form via fax to TECHNICAL SOLUTIONS CENTER at (302) 631-7597. If you have any questions or need additional information, please contact your local Technical Support Provider or Distributor or call (508) 668-5000.
- Initiated
- 2012-02-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-10-17
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Columbia, China, Denmark, Egypt, and Hong Kong.
- Recall number
- Z-1268-2012
- Recall status
- Terminated
- CDRH record ID
- 107628
- FDA classified
- 2012-03-21
- Recall event number
- 61236