Z-1268-2018: c6000, Chemistry (Photometric, Discrete), for clinical use. - cobas e 601 module: Immunoassay Analyzer, C

Recall record Z-1268-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1268-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Mar 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.

Source facts

Action
Roche Diagnostics Corporation issued Urgent Medical Device Correction (UMDC) notices and Fax Forms to their customers dated 12/18/2017 via UPS Ground (receipt required). Customers are encouraged to do the following Mandatory Steps to improve the vertical alignment and handling of 13 mm sample tubes and false bottom tubes: FOR cobas e 411 & cobas e 601 analyzers and MODULAR ANALYTICS E 170 module: 1) Always use Roche rack cup adapters with 13 mm tubes 2) Ensure sample tubes are straight FOR cobas e 411 disk system: 1) Always use 13 mm Sample Disk Tube Adapters (SDTAs) with 13 mm sample tubes. 2) Ensure sample tubes are straight. 3) Ensure 13 mm SDTAs are not exposed to direct sunlight. 4) If a larger diameter tube is left inserted in a 13 mm SDTA for more than 3 days, the adapter may become distorted. Do not insert smaller diameter tubes into distorted adapter as they may not align correctly. **All customers are encouraged to DISCARD Analyzer Bulletin 10-009U3. Follow the care and handling instructions for using adapters as outlined in this Urgent Medical Device Correction (UMDC) and complete & return the UMDC fax form via fax to 1-877-271-1472
Initiated
2017-12-18
Root cause
Labeling False and Misleading
Terminated
2019-04-09
Distribution
US Nationwide
Recall number
Z-1268-2018
Recall status
Terminated
CDRH record ID
161278
FDA record created
2018-03-28
Recall event number
79074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗