Z-1269-2021: Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000V Catheterization Table CAT-880B Canon CATHETERI

Recall record Z-1269-2021 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1269-2021
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Mar 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

Source facts

Action
On 01/12/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via USPS return receipt mail to customers informing them that the catheterization tables may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose. Additionally, the manual operation of the tabletop longitudinal or lateral movement will become heavy or hard to move due to the loosening of the bolts. Customers are instructed/informed: a. Should the table top tilt during a procedure before the bolts are replaced and secured, to ensure the safety of the patient, please move the table top to the position of the patients foot and refrain from operating the tabletop as much as possible. Please contact your service representative. b. Share this information with all users and reviewing radiologist as well as clinical engineering, and the biomedical group at their facility. c. Complete and return the attached Response Form and fax it to the toll free number 877-349-3054, or return via email to RAffairs@us.medical.canon For any questions or assistance, contact Regulatory Affairs Manager at 800-421-1968 or the local service representative at 800-521-1968.
Initiated
2020-12-08
Root cause
Component design/selection
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.
Recall number
Z-1269-2021
Recall status
Open, Classified
CDRH record ID
185815
FDA record created
2021-03-17
Recall event number
87322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗