Z-1269-2022: Alcon Clareon IOL with AutonoMe Delivery System

Recall record Z-1269-2022 for Alcon Laboratories Ireland, Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1269-2022
Company in FDA record
Alcon Laboratories Ireland, Ltd
FDA record creation date
Jun 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to device (preloaded intraocular lens) mislabeling, associated with partial mixing of product during overlapping manufacturing runs, and attributed to improper manufacturing line clearance.

Source facts

Action
On 05/11/2022, Alcon Laboratories sent an "URGENT: VOLUNTARY MEDICAL DEVICE REMOVAL" Letter via FedEx overnight informing customer that the firm has initiated a removal of one specific lot (25238769) of the Clareon IOL with AutonoMe Delivery System (CNA0T0.175). Based on the firm's investigation, there is the potential for limited number of 24.0D lenses to be included within this lot of the 17.5D Clareon IOL AutonoMe Delivery System. Customers are asked to locate and return the Alcon 17.5D Clareon IOL AutonoMe Delivery System with lot 25238769. An Alcon associate will reach out to customers directly to arrange for the return and replacement of the affected products. For questions, contact Alcon Customer Service at 1-800-862-5266 or contact their Alcon Sales Representative.
Initiated
2022-05-11
Root cause
Under Investigation by firm
Terminated
2025-02-05
Distribution
U.S.: AL, CA, FL, GA, IA, IL, KY, LA, MI, MO, NY, OH,OK, TX, VA, and WI O.U.S.: Czech Republic, France, Germany, India, Slovakia, Spain, and Switzerland
Recall number
Z-1269-2022
Recall status
Terminated
CDRH record ID
193722
FDA record created
2022-06-17
Recall event number
90194

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗