Open original FDA record ↗Recall record
- FDA identifier
- Z-1269-2025
- Company in FDA record
- Kung Shin Plastics Co. Ltd.
- FDA classification date
- Mar 4, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
Source facts
- Action
- On January 9, 2025, the firm sent a recall notice to their sole distributor in the US, followed by a more detailed notice on January 10. Customers were instructed to locate and immediately discontinue use of affected products. Affected product should be quarantined and returned. If products of affected batches have been sold, the distributor should immediately notify the buyers to stop using them, quarantine them, and destroy them.
- Initiated
- 2025-01-09
- Root cause
- Process design
- Distribution
- US Nationwide distribution in the state of IL.
- Recall number
- Z-1269-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212373
- FDA classified
- 2025-03-04
- Recall event number
- 96224