Open original FDA record ↗Recall record
- FDA identifier
- Z-1270-2014
- Company in FDA record
- Lin-Zhi International Inc
- FDA classification date
- Mar 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).
Source facts
- Action
- Lin-Zhi International sent a Recall letter dated February 18, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Letters acknowledge that Lin Zhi was notified by the customer that the product could not be qualified and that this was confirmed by Lin-Zhi internal testing. Letters request that the distributor notify all customers of the situation and ask them to destroy any remaining product from the affected lots. the root cause is identified and letters also state that a corrective action has been implemented to prevent future problems. Questions may be directed to Lin Zhi International at 408-732-3856.
- Initiated
- 2014-02-18
- Root cause
- Packaging
- Terminated
- 2014-04-25
- Distribution
- US Distribution including CA.
- Recall number
- Z-1270-2014
- Recall status
- Terminated
- CDRH record ID
- 125829
- FDA classified
- 2014-03-21
- Recall event number
- 67607