Open original FDA record ↗Recall record
- FDA identifier
- Z-1270-2021
- Company in FDA record
- Canon Medical System, USA, INC.
- FDA record creation date
- Mar 17, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Source facts
- Action
- On 01/12/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via USPS return receipt mail to customers informing them that the catheterization tables may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose. Additionally, the manual operation of the tabletop longitudinal or lateral movement will become heavy or hard to move due to the loosening of the bolts. Customers are instructed/informed: a. Should the table top tilt during a procedure before the bolts are replaced and secured, to ensure the safety of the patient, please move the table top to the position of the patients foot and refrain from operating the tabletop as much as possible. Please contact your service representative. b. Share this information with all users and reviewing radiologist as well as clinical engineering, and the biomedical group at their facility. c. Complete and return the attached Response Form and fax it to the toll free number 877-349-3054, or return via email to RAffairs@us.medical.canon For any questions or assistance, contact Regulatory Affairs Manager at 800-421-1968 or the local service representative at 800-521-1968.
- Initiated
- 2020-12-08
- Root cause
- Component design/selection
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.
- Recall number
- Z-1270-2021
- Recall status
- Open, Classified
- CDRH record ID
- 185816
- FDA record created
- 2021-03-17
- Recall event number
- 87322