Z-1270-2025: Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;

Recall record Z-1270-2025 for Telcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1270-2025
Company in FDA record
Telcare, LLC
FDA classification date
Mar 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.

Source facts

Action
On January 30, 2025, an IMPORTANT Safety Notification message was pushed out to display on all meter devices. IMPORTANT Safety Notification: https://www.gobio.com/bgm - information on proper battery use. Mensaje importante sobre batera. What you should do: If you have a Philips or BioTel charger and your meter is working normally, you can keep using it. If you experience any of these issues, stop using your meter and contact Philips Customer Service at (888) 666-4530 (US) or (1) (978) 610-4688 (International): 1. The meter looks damaged or cracked. 2. The meter is swollen. 3. The meter is too hot to hold. 4. The meter won t hold a charge. 5. You don t have the Philips or BioTel charger. How to Charge your meter safely: 1. Only use the USB cable and power adapter that were included in your kit. They are blue and branded with Philips or BioTel. 2. Keep the meter and cables away from water. 3. Store the meter between 10C and 40C (50F and 104F). 4. Keep the meter and cables clean and dry. Make sure to follow these guidelines to prevent battery issues. What Philips is doing: Philips will update you on how to improve the battery usage as soon as possible. We also encourage you to perform a Meter Replacement Check to determine if your device is working properly. Safety and quality are our top priorities. For more information or support, call Philips Customer Service at (888) 666-4530 (US) or (1) (978) 610-4688 (International), Monday Friday from 9:00 AM to 9:00 PM Eastern Time, excluding major holidays. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax: 1. Complete and submit the report Online. 2. Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 (1-800-332-0178).
Initiated
2025-01-30
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-1270-2025
Recall status
Open, Classified
CDRH record ID
212384
FDA classified
2025-03-05
Recall event number
96230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗