Z-1271-2020: Philips HeartStart FRx AED, Model 861304, FRx codeRev: E.02.017

Recall record Z-1271-2020 for Philips North America LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1271-2020
Company in FDA record
Philips North America LLC
FDA record creation date
Feb 19, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.

Source facts

Action
On 09/11/2019, the firm sent an "Customer Information Medical Device Recall" notification to customers via Certified Mail informing them that certain models of their Automated External Defibrillator (AED) contain a Printed Circuit Assembly (PCA) manufactured by a supplier that did not follow specific procedures when repairing PCAs that initially failed required testing. After being repaired, these PCAs passed all testing at the supplier's facility prior to being assembled into AEDs. The Recalling Firm's quality standards require that the PCAs both be repaired according to specified procedures and pass testing. The Recalling Firm states that the serial number of the AED is printed on the label on the back of the device. The Recalling Firm will replace the affected devices at no charge and while awaiting for the replacement AED, the customer may continue to use the device. The Recalling Firm or distributors will contact customers to arrange for replacement and removal of affected devices. For any additional questions/information, consumers can contact the Recalling Firm at (800)263-3342 option 5 to speak with a AED Recall Team or email heaertstartaedaction@philips.com
Initiated
2019-09-10
Root cause
Process control
Terminated
2024-04-29
Distribution
HeartStart FRx: US - CA, CT, FL, IL, MO, NY,TN, TX and WI OUS - Australia, Brazil, Canada, Germany, Italy, Japan, South Korea, Norway, Switzerland and Taiwan. HeartStart HS1: US - CT, FL, IL, MI, MS, NJ, TN, TX, WA and WI OUS - Argentina, Australia, Canada, France, Germany, Hong Kong, Italy, Japan, Netherlands, Norway, Spain, Sweden, and Taiwan
Recall number
Z-1271-2020
Recall status
Terminated
CDRH record ID
179536
FDA record created
2020-02-19
Recall event number
84850

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗