Open original FDA record ↗Recall record
- FDA identifier
- Z-1272-2021
- Company in FDA record
- Accuray Incorporated
- FDA record creation date
- Mar 17, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
Source facts
- Action
- An URGENT FIELD SAFETY NOTICE will be delivered to customers beginning on 02/26/2021. Customers are asked to return a signed copy of the letter back to Accuray. Field Service Engineers will visit all impacted sites to perform a work instruction that includes the inspection of the set screws and verification of the integrity of the overall couch system, including the linkage arm and its connection to the roll motor. This work instruction will be performed on an annual basis. Customers may continue to use the STC and CyberKnife System. However, those who assist patients should provide extra care when individuals are getting on or off the STC.
- Initiated
- 2021-02-26
- Root cause
- Equipment maintenance
- Terminated
- 2022-06-27
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Recall number
- Z-1272-2021
- Recall status
- Terminated
- CDRH record ID
- 186033
- FDA record created
- 2021-03-17
- Recall event number
- 87439