Skip to content

Public FDA record

Z-1272-2021: CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurger

Recall record Z-1272-2021 for Accuray Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1272-2021
Company in FDA record
Accuray Incorporated
FDA record creation date
Mar 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

Source facts

Action
An URGENT FIELD SAFETY NOTICE will be delivered to customers beginning on 02/26/2021. Customers are asked to return a signed copy of the letter back to Accuray. Field Service Engineers will visit all impacted sites to perform a work instruction that includes the inspection of the set screws and verification of the integrity of the overall couch system, including the linkage arm and its connection to the roll motor. This work instruction will be performed on an annual basis. Customers may continue to use the STC and CyberKnife System. However, those who assist patients should provide extra care when individuals are getting on or off the STC.
Initiated
2021-02-26
Root cause
Equipment maintenance
Terminated
2022-06-27
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-1272-2021
Recall status
Terminated
CDRH record ID
186033
FDA record created
2021-03-17
Recall event number
87439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗