Skip to content

Public FDA record

Z-1272-2022: Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-

Recall record Z-1272-2022 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1272-2022
Company in FDA record
Smiths Medical ASD Inc.
FDA record creation date
Jun 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001

Source facts

Action
The firm disseminated an URGENT MEDICAL DEVICE CORRECTION letter on 04/18/2022. The letter provides an overview of eight issues. These issues may result in pump alarm failure, falsely detected alarm failure, termination of active infusion, a state where the smart battery cannot provide its status to the pump, incorrect depleted battery alarm with interrupted infusion, abnormal internal clock behavior, pump run without the ability to terminate infusion via the Stop or Power keys, Program Volume Delivered (PVD) reset to zero, display of a false Rate Below Recommended Minimum for Syringe Size alarm, Pressure Increasing alarms occurring earlier than expected or false alarms with or without an occlusion present, display an incorrect value for the time remaining during a Bolus Dose or Loading Dose infusion while correctly infusing, and/or pump not communicating with PharmGuard Server (PGS).
Initiated
2022-04-18
Root cause
Device Design
Distribution
Worldwide distribution. US nationwide including US Virgin Islands, Countries of: AE, AL, AT, AU, BE, BM, BN, BS, CA, CH, CL, CY, CZ, DE, DK, DO, ES, FI, FR, GB, HK, HU, ID, IE, IL, IN, IR, IT, KR, KW, LB, LK, MU, MX, MY, NL, NZ, OM, PA, PH, PK, PL, PT, QA, SA, SE, SG, TR, TT, VI, and VN.
Recall number
Z-1272-2022
Recall status
Open, Classified
CDRH record ID
192909
FDA record created
2022-06-21
Recall event number
89913

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.