Z-1272-2023: Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOW

Recall record Z-1272-2023 for Vapotherm. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1272-2023
Company in FDA record
Vapotherm
FDA classification date
Mar 17, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.

Source facts

Action
A Technical Bulletin dated 10/27/22 was distributed to customers. Vapotherm has learned of certain Low Flow Disposable Patient Circuit (DPC) lots exhibiting excessive condensation. Please find the affected DPC lots in the attached appendix. You have been identified as a customer that received some or all of these lots. This condensation can be mitigated by ensuring that the delivery tube is fully primed and warm before placing the therapy on the patient, per existing instructions for use. Always check that the delivery tube is warm across the entire length before placing the therapy on the patient. If you experience delivery tubes that are not circulating water or are cold, do not use. Please contact Vapotherm Technical Support to report the issue. For every incident reported, we will replace and ship out a new DPC as soon as possible at no cost to you. Thank you for your continued support of Vapotherm and providing your patients with optimal care. If you have any questions, please contact Vapotherm Technical Support: For US customers, ts@vtherm.com or 855 557 8276 For UK customers, ukinfo@vtherm.com or +44 0345 305 8098 For all other customers, please reach out to your Vapotherm distributor. An Urgent Field Safety Notice dated 12/16/22 was sent to customers in the UK. Vapotherm requests that you return any unused products of the lots detailed below. Unused products from the affected lots should be returned as described in the Required Actions section within the FSN. This product removal is limited to the product codes / lot numbers listed below. No other product codes or lot numbers are affected. Required Actions to be taken by the VAPOTHERM: 1.For any unused product of the affected lots, replace affected DPC lots. 2.Email Customer Acknowledgment Form to impacted customers. 3.Ensure completion of Customer Acknowledgement Form. Required Actions to be taken by the END USER/ HEALTHCARE ORGANIZATION: 1.Inspect your inventory, locate any unused units of the af
Initiated
2022-12-16
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide and the country of UK.
Recall number
Z-1272-2023
Recall status
Open, Classified
CDRH record ID
198232
FDA classified
2023-03-17
Recall event number
91579

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗