Open original FDA record ↗Recall record
- FDA identifier
- Z-1273-2020
- Company in FDA record
- Smith & Nephew, Inc.
- FDA record creation date
- Feb 19, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
Source facts
- Action
- Smith Nephew "Urgent Medical Device Recall Notice R-2019-22" letter dated December 30, 2019. Customers were advised to: 1.Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
- Initiated
- 2019-12-30
- Root cause
- Package design/selection
- Terminated
- 2021-05-11
- Distribution
- Worldwide distribution US Nationwide distribution in the states of AZ, ID, MO, NC, ND, NE, OK, SC, TN, and countries of AE, AU, CA, CH, ES, FI, JP, PR, SG.
- Recall number
- Z-1273-2020
- Recall status
- Terminated
- CDRH record ID
- 179255
- FDA record created
- 2020-02-19
- Recall event number
- 84748