Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System.
Source facts
Action
In February 2021, Siemens will issue an Urgent Medical Device Correction notice to customers via letter stating that Siemens Healthcare Diagnostics Inc. has confirmed that the ADVIA Centaur HSV2 assay does not meet the 60 day onboard stability or the 28 day calibration frequency (interval) as stated in the ADVIA Centaur XP/XPT Herpes-2 IgG (HSV2).
Initiated
2021-02-08
Root cause
Under Investigation by firm
Terminated
2025-02-07
Distribution
Domestic: AL, CA, FL, IL, MD, NJ, NY, TN, TX. Foreign: Nepal, Chile, Georgia, Spain.
Recall number
Z-1274-2021
Recall status
Terminated
CDRH record ID
185931
FDA classified
2021-03-18
Recall event number
87381