Open original FDA record ↗Recall record
- FDA identifier
- Z-1274-2022
- Company in FDA record
- Olympus Corporation of the Americas
- FDA record creation date
- Jun 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
Source facts
- Action
- Olympus issued letter via e-mail on April 22, 2022. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide a service loaner device in an effort to not disrupt patient care. Contact: 647-999-3203 or at Cynthia.Ow@Olympus.com
- Initiated
- 2022-04-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-03-13
- Distribution
- US Nationwide distribution in the state of KY.
- Recall number
- Z-1274-2022
- Recall status
- Terminated
- CDRH record ID
- 193842
- FDA record created
- 2022-06-21
- Recall event number
- 90218