Skip to content

Public FDA record

Z-1274-2022: Colonovideoscope Model CF-H180AL

Recall record Z-1274-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1274-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Jun 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding

Source facts

Action
Olympus issued letter via e-mail on April 22, 2022. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide a service loaner device in an effort to not disrupt patient care. Contact: 647-999-3203 or at Cynthia.Ow@Olympus.com
Initiated
2022-04-22
Root cause
Nonconforming Material/Component
Terminated
2024-03-13
Distribution
US Nationwide distribution in the state of KY.
Recall number
Z-1274-2022
Recall status
Terminated
CDRH record ID
193842
FDA record created
2022-06-21
Recall event number
90218

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗