Z-1274-2023: FreeStyle Libre Reader, REF: 71936-01, 71937-01, 71938-01 a component of the FreeStyle Libre 14 Day Flash

Recall record Z-1274-2023 for Abbott Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1274-2023
Company in FDA record
Abbott Diabetes Care, Inc.
FDA classification date
Apr 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Source facts

Action
On 02/13/2023, correction notices were mailed or emailed to customers, health care professionals, and distribution partners who were asked to do the following: To determine if the Reader needs to be replaced, answer these questions: Does Reader appear damaged, cracked, or swollen? Has Reader been too hot to hold? Is it no longer able to hold a charge? If NO to all, perform a Reader Test. See www.FreeStyleBattery.com If yes or if the test failed, discontinue use and call Customer Service for replacement. - Don't expose the Reader, power adapter or yellow USB cable to liquids. - Store the Reader between -4 F and 140 F. - Avoid getting dust, dirt, blood, control solution, water, bleach, or any other substance in the test strip or USB port. - Check USB port and make sure it is dry and free of debris. - Use only the Abbott provided power adapter and yellow USB cable that came with your Reader and check they are free from damage. - Ensure you can access the power adapter and quickly disconnect it to prevent the potential risk of electrical shock. - Don't force or bend the end of the yellow USB cable into the Reader's USB port. - If the Reader is dropped or subjected to impact, perform a Reader Test to confirm it is still working properly. Customers are asked to acknowledge they have read and understood the actions contained in the letter by clicking the link or scanning the QR code. Health Care Professions were asked to ask their patients the questions above. Distribution Partners were asked to forward the correction notice and response form to customers. Customers with questions can contact Customer Service at 1-855-632-8658, available 7 days a week from 8AM to 8PM Eastern Time excluding major holidays.
Initiated
2023-02-13
Root cause
Device Design
Distribution
U.S. Nationwide.
Recall number
Z-1274-2023
Recall status
Open, Classified
CDRH record ID
198773
FDA classified
2023-04-06
Recall event number
91756

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗