Open original FDA record ↗Recall record
- FDA identifier
- Z-1275-2013
- Company in FDA record
- Cook, Inc.
- FDA classification date
- May 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation
Source facts
- Action
- Cook Medical sent an Urgent Medical Device Recall letter dated April 24, 2013 to all affect customers. The letter identified the affected products, description of the problem, and actions to be taken. The letter instructed customers to stop use, quarantine, and return all affected products. Customers were asked to complete product reply form and return affected products for credit to Cook Medical, ATTN: Retrun Goods/RGA # 2013C0003. For questions contact Cook Medical Customer Relations at 1-800-457-4500 or 1-812-339-2235.
- Initiated
- 2013-04-18
- Root cause
- Process control
- Terminated
- 2013-12-05
- Distribution
- Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
- Recall number
- Z-1275-2013
- Recall status
- Terminated
- CDRH record ID
- 117653
- FDA classified
- 2013-05-15
- Recall event number
- 64975