Z-1275-2013: Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver P

Recall record Z-1275-2013 for Cook, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1275-2013
Company in FDA record
Cook, Inc.
FDA classification date
May 15, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

Source facts

Action
Cook Medical sent an Urgent Medical Device Recall letter dated April 24, 2013 to all affect customers. The letter identified the affected products, description of the problem, and actions to be taken. The letter instructed customers to stop use, quarantine, and return all affected products. Customers were asked to complete product reply form and return affected products for credit to Cook Medical, ATTN: Retrun Goods/RGA # 2013C0003. For questions contact Cook Medical Customer Relations at 1-800-457-4500 or 1-812-339-2235.
Initiated
2013-04-18
Root cause
Process control
Terminated
2013-12-05
Distribution
Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
Recall number
Z-1275-2013
Recall status
Terminated
CDRH record ID
117653
FDA classified
2013-05-15
Recall event number
64975

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗