Skip to content

Public FDA record

Z-1275-2014: DICOM Export Format for Sum Dose with Velocity v3.0.0. A stand-alone software product that provides the

Recall record Z-1275-2014 for Velocity Medical Solutions, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1275-2014
Company in FDA record
Velocity Medical Solutions, LLC
FDA classification date
Mar 25, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An anomaly was detected such that, under certain conditions, a summed dose created by Velocity replaces the calculated dose from an approved plan in Eclipse when importing the dose from Velocity.

Source facts

Action
Velocity Medical Solutions sent a Urgent Field Safety Notice dated July 25th, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Velocity advises customers creating Dose volumes in Velocity to refrain from exporting these dose files to Eclipse until the customers install has been upgraded to Velocity 3.0.1, which will be available August 5th, 2013. A new version of Velocity that corrects this error, Velocity v3.0.1, will be available on August 5th, 2013 for upgrade at no cost. Please contact your Velocity Medical representative to obtain this upgrade. Service representative will upgrade the application at your earliest convenience. Please advise the appropriate personnel working in your medical department of the content of this letter. We sincerely apologize for any inconvenience and thank you in advance for your co-operation. Contact Information. Helpdesk: (404) 920-1966 or (855) 857-0802 and press 2.
Initiated
2013-07-25
Root cause
Software design
Terminated
2014-09-11
Distribution
Worldwide Distribution - US (nationwide) including the states of AL, CA, CO, CT, DC, FL, GA, IL, IN, LA, MA, MD, MI, MS, NC, NY, PA, SC, TN, TX, WI, and the countries of Belarus, Italy and Switzerland.
Recall number
Z-1275-2014
Recall status
Terminated
CDRH record ID
125842
FDA classified
2014-03-25
Recall event number
67615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗