Open original FDA record ↗Recall record
- FDA identifier
- Z-1275-2016
- Company in FDA record
- Vascular Solutions, Inc.
- FDA classification date
- Apr 7, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.
Source facts
- Action
- Consignees were sent on 3/4/2016 via first class mail a Vascular Solutions "Urgent Medical Device Recall" letter dated March 3, 2016. The letter described the problem and the product involved in the recall. It included the "Immediate Action Required of Customer / User" which instructed the customer to identify the product, remove it from inventory, place it in a secure area, and complete and return the VSI Account Inventory Form to the Customer Service Department.
- Initiated
- 2016-03-04
- Root cause
- Employee error
- Terminated
- 2016-10-19
- Distribution
- Nationwide Distribution in the states of AL, AR, CA, CO, DE, FL, GA, ID, IL IN, IA,KY, LA, MD, MA, MI, MO, NE,NV, NH, NY, NC, OH, PA, SD, TN, TX, UT, VA, WA, and WV. OUS: Ireland
- Recall number
- Z-1275-2016
- Recall status
- Terminated
- CDRH record ID
- 144454
- FDA classified
- 2016-04-07
- Recall event number
- 73553