Z-1275-2022: EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including t

Recall record Z-1275-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1275-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Jun 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications

Source facts

Action
Olympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Initiated
2022-04-22
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of AL, FL, NC, NY.
Recall number
Z-1275-2022
Recall status
Open, Classified
CDRH record ID
193859
FDA record created
2022-06-21
Recall event number
90232

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗