Open original FDA record ↗Recall record
- FDA identifier
- Z-1275-2022
- Company in FDA record
- Olympus Corporation of the Americas
- FDA record creation date
- Jun 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
Source facts
- Action
- Olympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
- Initiated
- 2022-04-22
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution in the states of AL, FL, NC, NY.
- Recall number
- Z-1275-2022
- Recall status
- Open, Classified
- CDRH record ID
- 193859
- FDA record created
- 2022-06-21
- Recall event number
- 90232