Z-1277-2018: CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and

Recall record Z-1277-2018 for Halyard Health, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1277-2018
Company in FDA record
Halyard Health, Inc
FDA record creation date
Mar 30, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.

Source facts

Action
The firm, Halyard, sent an "URGENT: Product Advisory Notice" letter dated 12/13/2017 to its customers on about 12/13/2017. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: Instructions for distributors included to examine unsold inventory for impacted lots, complete and return the Distributor Product Advisory Notice Acknowledgement Form, and send the attached End-User/Customer Product Advisory Notice and Acknowledgement Form to all end user customers who would have potentially received impacted product by email to: monica.king@hyh.com or by FAX to: 678-254-0347. A Halyard Customer Service Representative will contact the distributors if affected product was on hand. Instructions for End-User/Customers included to evaluate the clinical facility's unused inventory for impacted lots, inform all clinicians in the facility who may use potentially impacted devices to verify that the stylet tip is not visible within the window of the exit port after engaging the stylet with the port connector of the NG/NI Feeding Tube, and if it is visible, do not use the device. Additionally they were asked to complete and return the Product Advisory Notice Acknowledgement Form by email to: monica.king@hyh.com or by FAX to: 678-254-0347. If you require further assistance, please call Halyard Regulatory Affairs at 678-477-4165. You may also contact Halyard Customer Service at 1-844-HALYARD (1-844-425-9273, OPTION 1, OPTION 3).
Initiated
2017-12-13
Root cause
Under Investigation by firm
Terminated
2021-11-04
Distribution
Worldwide Distribution: US (nationwide) to states of: AR, CA, CO, GA, IA, LA, MI, MN, NM, TX; and country of Germany. Distributed internationally to Germany.
Recall number
Z-1277-2018
Recall status
Terminated
CDRH record ID
162433
FDA record created
2018-03-30
Recall event number
79479

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗