Open original FDA record ↗Recall record
- FDA identifier
- Z-1278-2012
- Company in FDA record
- Permobil Inc
- FDA classification date
- Mar 22, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The fasteners used between June 3, 2011 and January 31, 2012 to attach the Permolock C3 pins to wheelchairs manufactured in the United States do not contain the required nylon ring, which allows for a more secure attachment.
Source facts
- Action
- The firm, Permobil, sent an "URGENT DEVICE RECALL" letter dated February 27, 2012 via fax to all customers. For those customers where Permobil did not have a valid fax number, the notices were mailed via first class United States mail. The letter describes the product, problem and actions to be taken. Permobil will designate Permobil personnel to complete an update to the Permolock C3 device. If a Permobil representative has not contacted you already, please call 800-736-0925 to schedule the service.
- Initiated
- 2012-02-21
- Root cause
- Mixed-up of materials/components
- Terminated
- 2013-08-13
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, MD ,MI, MN, MO, NC, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WV, and WY; and countries of: Australia, Canada, France, Netherlands, Sweden, and United Kingdom.
- Recall number
- Z-1278-2012
- Recall status
- Terminated
- CDRH record ID
- 107875
- FDA classified
- 2012-03-22
- Recall event number
- 61351