Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip.
Source facts
Action
Reportable Event - Removal letters dated October 6, 2023 were sent to customers. Customers were instructed to return all product subject to recall and to notify any of their customers as well. Please return items for the attention of one of the PRRCs provided to ensure quick processing and assignment to the case. If you have any questions, please contact Mr. Sven Lazic (info@lazic.de +49 (0) 7461 96643 - 20) or Ms. Kim Maier (k.maier@lazic.de or +49 (0) 7461 96643 - 21).
Initiated
2023-10-06
Root cause
Device Design
Distribution
US Nationwide distribution in the state of ME.
Recall number
Z-1278-2024
Recall status
Open, Classified
CDRH record ID
205971
FDA classified
2024-03-11
Recall event number
93938
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.