Skip to content

Public FDA record

Z-1279-2016: Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Thi

Recall record Z-1279-2016 for Philips Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1279-2016
Company in FDA record
Philips Medical Systems, Inc.
FDA record creation date
Mar 31, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Loss of key image functionality due to a bent pedal of the Footswitch.

Source facts

Action
Field Safety Noticed was mailed on September 24, 2014 via Certified Return Receipt. This FCC consists of a Mechanical Hardware update of the Footswitch in order to prevent bent pedals and therefore prevents an intermittent or continuous inability of making live fluoro images or exposures. Customers with questions were instructed to contact their local Philips representative.
Initiated
2014-09-24
Root cause
Device Design
Terminated
2019-03-26
Distribution
Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC
Recall number
Z-1279-2016
Recall status
Terminated
CDRH record ID
130261
FDA record created
2016-03-31
Recall event number
69339

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗