Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
Source facts
Action
A customer notification letter will be sent to affected customers by email. All the affected systems will be corrected with the upgraded software in the field free of charge.
Initiated
2018-10-22
Root cause
Software design
Terminated
2022-08-25
Distribution
US Nationwide Distribution.
Recall number
Z-1279-2019
Recall status
Terminated
CDRH record ID
172628
FDA record created
2019-05-21
Recall event number
82753