Z-1280-2023: Scan Monitor/ScanWatch, model number hwa09 Wearable, Bluetooth-connected wrist watch that records heart

Recall record Z-1280-2023 for Withings. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1280-2023
Company in FDA record
Withings
FDA classification date
Mar 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software bug eliminated the initial ECG activation and review of the ECG under a physician. However, the device is only cleared for use under the care of a physician.

Source facts

Action
Users were initially notified via push notifications in the companion app on December 19, 2022, which instructed users to update the app and activate the ECG feature under a physician. On February 10, 2023, the firm followed up with customers via Medical Device Recall letters. Customers were instructed to ensure their Health Mate app is updated to iOS 5.13.0 or higher, and to ensure their device is synchronized with the Health Mate app. Customers should follow the instructions on their Health Mate app to activate their ECG feature. Contact Information for questions: Email: deals-support@withings.com Phone number: 1-888-229-4116 (Availability Mon - Fri, 08:00 - 22:00 EST, 13:00 - 03:00 GMT, 14:00 - 04:00 CET).
Initiated
2022-12-19
Root cause
Software Design Change
Distribution
Nationwide distribution in the US. No foreign distribution.
Recall number
Z-1280-2023
Recall status
Open, Classified
CDRH record ID
198114
FDA classified
2023-03-22
Recall event number
91523

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗