Skip to content

Public FDA record

Z-1281-03: Stanmore modular hip system; Stanmore CoCr femoral size 1 std. stem; part 164241

Recall record Z-1281-03 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1281-03
Company in FDA record
Biomet, Inc.
FDA classification date
Sep 24, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The labeling does not instruct the removal of the protective shipping cap, and one surgeon implanted with the cap still in place.

Source facts

Action
Distributors were sent a letter on June 18, 2003 via FedEx. This mailing contained labels and the distributors were instructed to affix the label to the inventory consigned to hospitals within their territory.
Initiated
2003-06-18
Root cause
Other
Terminated
2003-09-23
Distribution
United States
Recall number
Z-1281-03
Recall status
Terminated
CDRH record ID
27999
FDA classified
2003-09-24
Recall event number
26618

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.