Open original FDA record ↗Recall record
- FDA identifier
- Z-1281-2015
- Company in FDA record
- Synthes, Inc.
- FDA classification date
- Mar 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Source facts
- Action
- An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: Synthes6734@stericycle.com Option 2: Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility. Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
- Initiated
- 2015-02-17
- Root cause
- Process control
- Terminated
- 2016-08-26
- Distribution
- Nationwide Distribution
- Recall number
- Z-1281-2015
- Recall status
- Terminated
- CDRH record ID
- 133826
- FDA classified
- 2015-03-16
- Recall event number
- 70455