Skip to content

Public FDA record

Z-1281-2015: Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.0

Recall record Z-1281-2015 for Synthes, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1281-2015
Company in FDA record
Synthes, Inc.
FDA classification date
Mar 16, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.

Source facts

Action
An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: Synthes6734@stericycle.com Option 2: Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility. Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
Initiated
2015-02-17
Root cause
Process control
Terminated
2016-08-26
Distribution
Nationwide Distribution
Recall number
Z-1281-2015
Recall status
Terminated
CDRH record ID
133826
FDA classified
2015-03-16
Recall event number
70455

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.