Open original FDA record ↗Recall record
- FDA identifier
- Z-1281-2021
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA record creation date
- Mar 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or later
Source facts
- Action
- Raysearch issued via E-mail on 03 March 2021, Field Safety Notice, Medical Device Correction #75198 letter stating reason for recall, health risk and action to take: RaySearch will check if the issue affects your system and apply the below solution. " Until the solution has been applied, be aware that the message body is missing from messages created in RayCare 2. " Confirm you have read and understood this notice by replying to the notification email. For regulatory information, please contact quality@raysearchlabs.com SOLUTION This issue will be resolved by RaySearch staff using a support tool. Using the support tool, the message body will be restored to all messages in the currently used RayCare version.
- Initiated
- 2021-03-03
- Root cause
- Software design
- Distribution
- US Nationwide distribution in the state of TN.
- Recall number
- Z-1281-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186123
- FDA record created
- 2021-03-19
- Recall event number
- 87471