Open original FDA record ↗Recall record
- FDA identifier
- Z-1281-2023
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA classification date
- Mar 22, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
Source facts
- Action
- A Field Safety Notice, Medical Device Correction letter dated 2/16/23 was sent to customers. Actions to be taken by the user " Be aware that an RT plan exported by using the VSIM Export button will always have the DICOM attribute Beam Limiting Device Angle (300A,0120) set to zero. " Be aware that an RT plan exported by using the VSIM Export button will always have the DICOM attribute Number of Wedges (300A,00D0) set to zero and no wedge information will be exported. " Be aware that an RT plan exported by using the VSIM Export button will never include the DICOM attribute Applicator Sequence (300A,0107) and no cone information will be exported. " Do not use or export Virtual Simulation plans with non-zero collimator angle, wedges or cones using the VSIM Export button. " Be aware that a plan imported as a Virtual Simulation plan may not contain all intended information and that some attributes affecting dose may be different from the original plan. " Always perform dose computation for imported Virtual Simulation plans before export. Do not use the VSIM Export for export of such plans. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. Solution This issue will be resolved in the next version of RayStation, scheduled for market release in June 2023 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. Transmission of this Notice This notice needs to be passed on to all those who need to be aware within your organization. Maintain awareness of this notice as long as any affected version is in use.
- Initiated
- 2023-02-17
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide
- Recall number
- Z-1281-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199077
- FDA classified
- 2023-03-22
- Recall event number
- 91867